Sign in

User name:(required)

Password:(required)

Join Us

join us

Your Name:(required)

Your Email:(required)

Your Message :

0/2000

Types and Usage Details of Endoscopic Biopsy Forceps

Author: Molly

Jun. 17, 2024

Types and Usage Details of Endoscopic Biopsy Forceps

There are many types of endoscopic biopsy forceps


Are you interested in learning more about Disposable Biopsy Forceps? Contact us today to secure an expert consultation!

Endoscopic biopsy forceps include standard type, side-opening type, head-end needle type, circular cup-shaped type, tooth-shaped cup-shaped type, and so on. They are mainly connected by laser welding, which can be achieved by continuous or pulsed laser beams. Laser radiation heats the surface to be processed, and the surface heat spreads inward through heat conduction. By controlling laser parameters such as pulse width, energy, peak power, and repetition frequency, the workpiece is melted to form a specific pool.


The energy conversion mechanism is completed through the "small hole" structure. Under sufficient power density laser irradiation, the material evaporates and forms a small hole. This vapor-filled small hole is like a black body that absorbs almost all of the incident light beam energy. The equilibrium temperature inside the hole is about °C, and the heat is transferred from the high-temperature hole wall to melt the metal surrounding the hole cavity. The high-temperature steam continuously evaporated by the wall material under the irradiation of the light beam fills the small hole, and the wall surrounding the small hole is surrounded by melted metal, and the liquid metal is surrounded by solid material. The flowing liquid outside the hole wall and the surface tension of the wall layer are balanced with the continuous vapor pressure generated inside the hole cavity, maintaining dynamic equilibrium. The light beam continuously enters the small hole, and the material outside the small hole flows continuously. The small hole is always in a flow-stable state as the leading light beam moves forward at a certain speed. That is to say, the small hole and the melted metal surrounding the wall of the hole move forward with the leading light beam, and the melted metal fills the gap left after the small hole moves away and then condenses, forming a weld.


All of the above processes happen so quickly that the welding speed can easily reach several meters per minute. This is the mechanism we see for the formation of screw cavities of biopsy forceps. Therefore, once the thread cavity of the biopsy forceps is broken, it cannot be repaired by ordinary welding, and metal burrs will be formed. In recent years, most biopsy forceps have adopted a rigid four-bar linkage structure, making the use of biopsy forceps more convenient. Here we pay tribute to the tireless engineers, who have devoted a lot of effort from invention to market application of a product.


Details of using biopsy forceps


The key to using biopsy forceps is to master the strength of your hands. I am used to using the ring finger, middle finger, index finger, and thumb to cooperate with the use of biopsy forceps. When using endoscopic biopsy forceps, the force should be moderate, and avoid using excessive force to switch. This not only cannot grasp the diseased tissue well but also easily damages the biopsy forceps.


The use of biopsy forceps in mastering strength is the foundation of various accessories. You may not feel it during ordinary biopsy, but when you are removing foreign bodies, especially coins, when using alligator forceps to pick up coins, if the alligator forceps are opened too much and used too much force, it is difficult to firmly clamp the coin. Open the forceps slightly larger than the edge of the coin, and try to go over the edge of the coin. Then clamp the coin firmly with force. Pay attention to practicing opening the biopsy forceps to any size and self-feeling.

Performance of disposable endoscopic forceps according ...

In the in vitro model, the molding forceps provided a significantly higher popping rate than the other forceps (cutting forceps, 25.0%; pressing forceps, 17.5%; and molding forceps, 41.7%; p = 0.006). In the in vivo model, the cutting and pressing forceps did not provide larger specimens, deeper biopsy specimen, and higher specimen adequacy than those obtained using the molding forceps (p = 0., p = 0., and p = 0., respectively). However, the molding forceps showed significantly more common crush artifact than the others (cutting forceps, 0%; pressing forceps, 5.0%; and molding forceps, 43.3%; p = 0.).

By using an in vitro nitrile glove popping model, we compared the popping ability among eight different disposable biopsy forceps (one pair of biopsy forceps with cups made by a cutting method [cutting forceps], four pairs of biopsy forceps with cups made by a pressing method [pressing forceps], and three pairs of biopsy forceps with cups made using a injection molding method [molding forceps]). Using an in vivo swine model, we compared the penetration depth and quality of specimen among the biopsy forceps.

The disposable biopsy forceps do not have more functional problems than the reusable ones, but they do have various types of technical issues affecting cost-effectiveness. To reduce the production costs, either less costly methods could be used instead of precision manufacturing methods for reusable biopsy forceps, or low-priced materials instead of main materials could be used. The most essential problem in terms of the technical aspects or production costs is the cup portion that catches and obtains specimens. As one of the manufacturing methods of this cup portion, the injection molding method has become widely used. It has a very short processing time and a capability of mass production, with lead to cheap prices. However, these low-priced disposable biopsy forceps are currently being widely used without a proper evaluation of the quality of specimens obtained, biopsy performance, and complication rate in comparison with the other types of forceps.

Meanwhile, with reports indicating cross-contamination and spread of infection by endoscopic accessories, the reusable biopsy forceps have been replaced by the disposable ones [ 6 - 8 ]. However, the replacement process was not easy because of the fixed insurance payment system in some countries. There have been even some cases where disposable biopsy forceps were reprocessed to be reused several times as an interim measure. The biggest technical issue regarding reusing disposable biopsy forceps is that it can cause deformation of the forceps, not to mention cross-contamination. Unlike reusable biopsy forceps, which are made to endure exposure to sterilants or disinfectants, disposable biopsy forceps are easily changed or deformed when damaged by reprocessing. As such, the working channel of the endoscope can be damaged, which would incur high repair costs or various unexpected problems in biopsy sampling. To resolve these issues, disposable biopsy forceps with lower production costs were required to overcome the limitations of the insurance payment system. Many countries, including the United States, used this method until , when the change was triggered by the U.S. Food and Drug Administration during the presentation of a guideline to regulate single-use devices. Currently, professional companies are commissioned to perform a more thorough reprocessing, or cheaper disposable biopsy forceps are used only once and then discarded [ 9 , 10 ].

In the field of gastrointestinal endoscopy, the development of new devices and related programs has led to the advancement of diagnostic methods at an exponential rate, but direct histopathological examination is still the most important and accurate diagnostic method. Therefore, the quality of endoscopic biopsy specimen is crucial for a successful histological diagnosis. The basic concept of biopsy forceps has not changed much, but various types of biopsy forceps with different designs and materials are widely used [ 1 ]. Previous reports indicated that the quality of biopsy specimens depends on the shape and size of the forceps cups [ 2 - 5 ]. However, as endoscopes are becoming slimmer, smaller-sized cups have become more widely used, since larger-sized cups limit the diameter of the working channel. This resulted in technological advancements, including changing the shape of cups from oval to elongated shape, processing the edges of cups in a serrated or toothed-jaw cutting shape, or inserting a stabilizing needle or spike between cups, using more precise processing methods [ 1 ].

Contact us to discuss your requirements of Laparoscopic Instrument. Our experienced sales team can help you identify the options that best suit your needs.

The mean ± standard deviation (SD), and range were used to summarize the data for continuous variables and percentages for categorical variables. The Kruskal-Wallis test was used to analyze the difference of specimens between the groups. A p value of < 0.05 was considered significant. Statistical analysis was performed with the IBM SPSS Statistics version 21.0.0 software (IBM Co., Armonk, NY, USA).

The same technique was used for the preparation of the histology slides for tissues obtained by all forceps. The best specimen from each set was used for statistical analysis. Specimens were fixed with 10% neutral-buffered formalin and processed routinely in a Thermo Shandon Exelsior ES tissue processor (Thermo Fisher Scientific, Waltham, MI, USA). Paraffin-embedded tissues were sectioned on a microtome at 4 μm and mounted on the glass slides for hematoxylin and eosin staining.

All biopsy specimens were examined by an experienced gastrointestinal pathologist (G.K.) who was blinded to the type of the biopsy forceps tested. Each specimen was analyzed for maximal diameter (measured in millimeter), depth of acquired specimen (lamina propria, muscularis mucosa, or submucosa), adequacy for diagnosis (adequate or inadequate; general adequacy of the specimens for diagnosis by the experienced gastrointestinal pathologist), and crush artifact (none, mild, moderate, or severe; severe crush artifact can affect the diagnosis) ( ) [ 5 ].

Using a forward endoscope (GIF-Q260J, Olympus Co.), an expert endoscopist (W.J.K.) who was blinded to the type of the biopsy forceps performed the endoscopic examination with biopsy sampling. Using two pairs of biopsy forceps of each type, five specimens were obtained from the greater curvature side of the antrum in one swine and five specimens from the opposite lesser curvature side of the antrum in the other swine ( ).

Two mini-pigs (Sus scrofa; mean age, 14 months; mean body weight, 30.2 kg) were used for the in vivo swine study. Before this study, the pigs were fasted for > 24 hours, except for water given ad libitum. The pigs were pre-sedated using an intramuscular injection of atropine sulfate (0.04 mg/kg), xylazine (2 mg/kg), and tiletamine-zolazepam (5 mg/kg). Then, the pigs were intubated, and general anesthesia was achieved using 0.5% to 2% isoflurane through the endotracheal tube with 70% nitrous oxide and 30% oxygen provided by a ventilator. For endoscopic examination, the pigs were placed on their left lateral sides on a table.

This is a commonly used test to determine the grasping ability of biopsy forceps. The tips of air-filled nitrile gloves (Nitrisoft, Kossan Latex Industries Sdn. Bhd., Klang, Malaysia), which are used during endoscopic examination, were grasped and pulled with biopsy forceps to determine if there was any cutting or tearing of the gloves ( ). Cutting and tearing are not generally distinguished when assessing grasping ability, but the cutting phenomenon refers to popping with a sharp cut of the part grasped inside the cups of the biopsy forceps, while tearing phenomenon refers to popping with a large tear of the surrounding areas and not just a cut of the part grasped inside the cups. These two are not accurately distinguishable, and a better test result is not necessarily indicative of a better grasping ability, which is one weakness of the test. Furthermore, cups with smoothly processed surface may be slippery when attempting to grasp the gloves and thus may not create any popping. This is a very simple, widely used method, but determining whether it can successfully access the performance of biopsy forceps is highly important. Thus, this test was included in this study to determine its objective accuracy. Using four biopsy forceps of each type, 20 popping tests, five times for each forceps, were performed. The biopsy forceps were tested in a random order by a tester (W.J.K.) who was blinded to the type of the biopsy forceps.

Generally, the cost of producing the cups of the cutting forceps is approximately 10 times more than that of the other two methods. The costs per unit of the cups of the pressing and molding forceps are almost same. However, the two methods require mold making, which costs twice more for the pressing forceps than for the molding forceps. In addition, in the lifetime of the mold of the pressing forceps, bending or the folding thin metal boards can easily induce stress to the mold. Therefore, pressing forceps require more development/higher production costs because more-frequent mold replacement is needed than that for the molding forceps.

The manufacturing techniques for the cup portion of disposable biopsy forceps are classified into three types as shown in . First, the cutting method using precision lathe technique (cutting forceps hereinafter) uses a lathe to precisely cut or process metals. For this reason, the cups have the smoothest surface, and good finishing make them the most superior in terms of tolerance factor. In addition, their design can be changed freely, facilitating improvement. However, the cutting forceps have the highest processing and production costs, and the lowest productivity among the three methods. Second, the pressing method used the press molding technique (pressing forceps hereinafter), which includes bending or folding thin metal boards. Compared with the cutting forceps, they have a superior productivity, lower production costs, and non-inferior tolerance factor. On the other hand, owing to the limitation in precise design, it has a lower degree of precision and makes a change in the design more difficult than that of the cutting method. Third, the cheapest method using the injection molding process (molding forceps hereinafter) makes a specific shape by applying heat to metal powder in a designated mold. Having the most superior productivity among the three methods, it can be mass-produced at a very low cost. However, it has a very coarse metal surface, crude design, and the possibility of having an ejector pin mark left while removing it from the mold. It also has a very high error rate in production and a very low degree of precision. A secondary precise surface finishing can be performed to obtain better quality, but not to avoid an increase in the production costs.

The summary of the results on the performance of different types of forceps according to the manufacturing techniques are shown in . In terms of specimen size, no significant difference was observed between the three types (cutting forceps, 3.70 ± 1.22 mm; pressing forceps, 4.15 ± 1.24 mm; and molding forceps, 3.79 ± 0.89 mm; p = 0.) ( ). In terms of depth involvement, no significant difference was observed between the three groups (cutting forceps, 30%; pressing forceps, 32.5%; and molding forceps, 43.3%; p = 0.). In terms of specimen adequacy, no significant difference was observed between the three types (cutting forceps, 90%; pressing forceps, 97.5%; and molding forceps, 100%; p = 0.). However, the molding forceps had significantly more common crush artifact than the other forceps (cutting forceps, 0%; pressing forceps, 5.0%; and molding forceps, 43.3%; p = 0.).

A summary of the performance of different types of forceps according to the producing company is shown in . When comparing the specimen size, no significant difference was observed between the eight types of forceps (p = 0.) ( ). In terms of the depth involvement, no significant difference was present (reaching % to the muscularis mucosa or submucosa; p = 0.). As for the specimen adequacy to enable examination for diagnosis, no significant difference was observed between the three types (% of adequate specimens for diagnosis; p = 0.). However, types D, E, and H had a significantly more common crush artifact than the others (p = 0.).

DISCUSSION

When disposable biopsy forceps share a similar shape and size of cups as those of reusable biopsy forceps, they are known to have similar performance in terms of specimen size, histological depth, and so forth [11]. The performance of biopsy forceps can vary depending on the manufacturing technique of cups, but simple tests such as the glove popping test have been commonly used to assess the performance of biopsy forceps. This is the first study to investigate if there is any difference in performance between the disposable biopsy forceps based on manufacturing techniques and to determine the accuracy of the glove popping test. We showed for the first time that a better result of the glove popping test does not necessarily mean better performance, and that crush artifact, but not specimen size and adequacy, can vary considerably depending on the manufacturing technique of cups.

As mentioned in the introduction section, accessories and endoscopes have had continuous and steady technological advancements. Further development has been restrained by the transition from reusable to disposable accessories. Biopsy forceps are the simplest among endoscopic accessories, with their basic function being specimen obtainment. Therefore, they have advanced in the direction of reducing production cost, rather than technological development. The transition from reusable to disposable biopsy forceps was also a very important issue for manufacturers, and the production cost could not be reduced without considering the performance of biopsy forceps. For this reason, making disposable biopsy forceps in the same way as reusable biopsy forceps, such as the cutting method using precision lathe technique, was highly inefficient in terms of production cost. For this reason, there was only one company that manufactured cutting forceps. Consequently, pressing and molding forceps were introduced sequentially to continue the effort to reduce the production costs. Many third-party companies and major companies in Korea continue to develop new disposable biopsy forceps, but most of them focus on lowpriced molding forceps. The reality is that the emphasis is predominantly placed on the user&#;s convenience and the design of the handle rather than the performance of cups. Although some institutions are using the glove popping test to assess the biopsy forceps objectively, the actual association between this test and the performance of forceps has not been reported yet.

The problem of the glove popping test, as mentioned earlier, is that the test result can be good not only when cutting is done accurately with good performance ( ) but also when the glove is torn due to the coarse surface of the cups ( ). As can be seen in , which shows the results of the glove popping test, the results of the molding forceps were generally good. However, these results bear little similarity when compared with the results of performance shown in and . The assumption that the tearing phenomenon by rough surface of cups would be the mechanism of glove popping by the molding forceps can be made based on the fact that crush artifact was more frequent than in the other groups. However, because of the very high popping rate in type D forceps, there was a significant bias that increased the popping rate of the entire molding forceps. Nevertheless, we think that evaluating the performance of biopsy forceps using the popping test is not accurate because we did not find any differences between the groups. We would like to argue that using the glove popping test alone is not recommended when assessing the performance of biopsy forceps.

Crush artifact, which does not occur when the specimen grasped inside the cups cuts cleanly, refers to a serious compression in the middle part when some of the tissues outside the cups are torn apart because of poor grip ability. As can be seen in , crush artifact can be graded depending on the extent of base tissue, which has not been grasped but has been pulled out without being cut cleanly from the tissue. A severe grade means that the compression is serious enough to affect the diagnosis [5]. As shown in , fortunately, none of the three groups had severe crush artifacts, but crush artifacts were most common with the molding forceps. This means that during an endoscopic biopsy, complications may occur depending on the type of biopsy forceps used and the condition of the base tissue. This also implies that when performing biopsy of deep ulcer or relatively thin duodenum or colon, the specimen may not cut cleanly and some of the base tissue can be torn, causing complications, such as bleeding or perforation. A previous report that only compared pressing forceps also reported a 6% to 7% incidence of crush artifact, which was similar to our results [5]. This report also explained that little crush artifact means lack of tissue compression with a good tissue fenestration due to good grip ability of the cups.

The limitations of this study are as follows: (1) in vitro and in vivo animal study; (2) small sample size of biopsy specimens per type of biopsy forceps; (3) performed gastric biopsies only; (4) assessing performance simply according to the manufacturing techniques of cups, given that other parts of biopsy forceps can also affect the performance; (5) lack of subtype analysis because not many types of cups were used; (6) lack of analysis of biopsy-induced performance complications such as immediate bleeding; and (7) the adequacy of biopsy specimen and crush artifact may depend on the pulling force of the endoscopist. The participation of two or more endoscopists is recommended for a well-designed future study.

In conclusion, this was the first study to compare the performance of biopsy forceps according to manufacturing techniques. We found that the molding forceps provided lower performance than the cutting and pressing forceps in terms of crush artifact. In addition, the simple glove popping test should not be recommended for the evaluation of performance of biopsy forceps.

If you are looking for more details, kindly visit Disposable Trocar.

66

0

Comments

0/2000

All Comments (0)

Guest Posts

If you are interested in sending in a Guest Blogger Submission,welcome to write for us!

Your Name:(required)

Your Email:(required)

Subject:

Your Message:(required)

0/2000